SCPI Command Reference 
97 
 
6.3.7    TRIGger Subsystem 
The Trigger subsystem is made up of a series of commands to configure the Trigger model. These 
commands and subsystems are summarized in Table 6-7 
Table 6-7 
TRIGger Command Summary 
Command 
Function Description 
Default   
:INITiate 
  [:IMMediate] 
  :CONTinuous <b> 
  :CONTinuous? 
 
:ABORt 
:TRIGger 
    :SOURce <name> 
    :SOURce? 
    :DELay <n> 
:AUTO <b> 
      :AUTO? 
Subsystem command path 
Initiate one trigger cycle. 
Enable or disable continuous initiation. 
Query continuous initiation. 
 
Reset trigger system 
 
Select control source 
Query control source 
Set delay time 
Enable or disable auto delay 
Query state of delay 
 
 
 
 
 
 
 
IMMediate 
 
OFF 
 
 
:INITiate 
[:IMMediate] 
Command Syntax: 
:INITiate[:IMMediate]   
 
Description: 
This command takes the instrument out of the idle state. After all programmed operations are 
completed, the instrument returns to the idle state if continuous initiation is disabled (see next 
command). 
 
 
:CONTinuous <b> 
Command Syntax: 
:INITiate:CONTinuous <b>  
 
Control continuous initiation 
 
Command parameter: 
<b> =    0 or OFF   
 
 
 
Disable continuous initiation 
1 or ON   
 
 
 
Enable continuous initiation 
Query: 
:CONTinuous?     
 
 
 
Query continuous initiation 
Return String: 
 
1 or 0 
 
Description: 
When continuous initiation is selected (ON), the instrument is taken out of the idle state. At the end