Gw Instek LCR-819 Digital-Multimeter, DMM, LCR-819 Hoja De Datos

Los códigos de productos
LCR-819
Descargar
Página de 91
 
18
z
  Continuous Command 
When you would like to send two commands or more, you have to send a code between two commands. Refer to below: 
<NL^END> or <^J> : New Line or ASCⅡ Line Feed Character (Hex 0A). 
Last Command need add two codes. Refer to below:   
<NL^END> or <^J> : New Line or ASCⅡ Line Feed Character (Hex 0A). 
<CR^M> :  ASCⅡ Carry Return Character (Hex 0D) 
Example: Setting Frequency, Voltage and Test Speed 
PC send command   
MAIN
:
FREQ
 
1.00000
<
 ^END
>(or 
<^J>)
 MAIN
:
VOLT
 
1.000
<
 ^END
>(or 
<^J>)
 MAIN
:
SPEE
:
FAST
<
 ^END^M
>(or 
<^J^M>)
 
ASCII CODE  = 
4D 41 49 4E
 3A 
46 52 45 51
 20
 31 2E 30 30 30 30 30
 0A 
4D 41 49 4E
 3A 
56 4F 4C 54
 20 
31 2E 30 30 30 
0A 
4D 41 49 
4E
 3A 
55 50 45 45
 3A 
46 41 53 54
 0A 0D  (Hex format)