   GETTING 
STARTED 
23 
 
 
+
Power
Supply
200mA
MAX
MAX
30V
CAT
EXTERNAL  BIAS
+
LCD
Don’t connect
 
 
 
   LCR-800 
User 
Manual 
24
 
Zeroing 
Zeroing calibration 
 
Background 
Open and short circuit calibration (zeroing) should 
be performed on a daily basis to correct for cable 
and fixture errors before taking measurements. 
When test fixtures or test cables are changed, the 
zeroing process should be performed again. All 
data performed during the calibration is stored in 
the internal memory of the LCR-800.  
The Open circuit calibration determines the stray 
admittance and compensates high impedance 
measurements. The short calibration determines 
the residual impedance and is used when 
determining low impedance measurements. 
 
Open circuit 
The Open circuit calibration measures the stray 
admittance of the test fixture. This is used for high 
impedance measurements. 
Procedure 
1.  Insert the test fixture or cable. 
Ensure the cables are not 
shorted and are open.  
 
 
 
2.  Press the MENU key, then OFFSET, followed 
by CAP OFFSET. 
 
 
 
 
3.  Wait for the calibration to finish. If the OPEN 
TEST was successful, the screen will display the 
following message: