Gw Instek LCR-816 Digital-Multimeter, DMM, LCR-816 Hoja De Datos

Los códigos de productos
LCR-816
Descargar
Página de 91
  
RS232 
REMOTE 
77 
 
 
Note 
DataBits, StopBits, Parity and Flowcontrol cannot be 
edited. 
Note 
All file menus (File, Option, Data, Help) are restricted 
in Auto mode. To change to manual mode see page 43 
or 80 to change to Manual mode manually or 
remotely. 
   LCR-800 
User 
Manual 
78
 
LCR Viewer File Settings 
 
Background  
The LCR Viewer file system stores 10000 test 
results per file. The files are comprised of the file 
name identifier and file number identifier. 
 
1  file name identifier 
2  file number identifier 
3  TXT File extension 
The file name identifier consists of 4 user-defined 
characters. The file number identifier is 
incremented per 10000 test results. If LCR Viewer 
is terminated before 10000 test results, the data will 
be saved and then the next file will start anew. The 
file number identifier starts at 0001 and increments 
to a maximum of 9999. The file number identifier 
cannot be user-defined, but can be reset to 0001. 
Test Result 
File Name 
Identifier 
File_Name 
File Number 
Identifier 
File_Num Filename 
1~10000 LCR_ 
0001 
LCR_0001.txt 
10001~20000 LCR_ 
0002 
LCR_0002.txt 
99980001~ 
99990000 
LCR_ 9999  LCR_9999.txt 
 
1.  Ensure the LCR-800 is set to 
manual (single) measurement 
mode. 
Page 80 
 
2.  Go to the OptionSettings menu.