PROGRAMMING 
99 
Example MAIN:EXTB:OFF.<^END^M> 
 
Turn External Bias off. 
Query Syntax 
MAIN:EXTB?<^END^M>or<^J^M> 
Return String 
<string> 
External Bias status 
MAIN:EXTB:OFF.<^END> 
MAIN:EXTB:ON..<^END> 
Off 
On 
Query Example 
MAIN:EXTB?<^END^M> 
MAIN:EXTB:ON..<^END> 
 
Returns the External Bias status (On). 
PPM 
Command/Query 
 
Turns PPM on or off for Dissipation (D) or Quality factor (Q) 
measurements.   
Syntax MAIN:PPM.:<string><^END^M>or<^J^M> 
Parameter 
<String>  
PPM 
 
OFF.  
Off 
ON..  
On 
Example MAIN:PPM.:OFF.<^END^M> 
 
Turns PPM off. 
Query Syntax 
MAIN:PPM.?<^END^M>or<^J^M> 
Return String 
<string> PPM 
status 
MAIN:PPM.:OFF.<^END> 
MAIN:PPM.:ON..<^END> 
Off 
On 
Query Example 
MAIN:PPM.?<^END^M> 
MAIN:PPM.:ON..<^END> 
 
Returns PPM status (On). 
   LCR-800 
User 
Manual 
100
 
OPEN 
Command 
 
This command will perform an open circuit calibration. A return 
string will indicate if the calibration was successful or not. 
Syntax OFFS:OPEN<^END^M>or<^J^M> 
Return String 
<string> 
Open calibration attempt 
OPEN:OK<^END> 
OPEN:FAIL<^END> 
Successful 
Failure 
Example OFFS:OPEN<^END^M> 
OPEN:OK<^END> 
 
Returns the open circuit calibration attempt 
(Successful). 
SHORT 
Command 
 
This command will perform a closed (short) circuit calibration. A 
return string will indicate if the calibration was successful or not. 
Syntax OFFS:SHOR<^END^M>or<^J^M> 
Return String 
<string> 
Short calibration attempt 
SHOR:OK<^END> 
SHOR:FAIL<^END> 
Successful 
Failure 
Example 
OFFS:SHOR<^END^M> 
SHOR:OK<^END> 
 
Returns the closed circuit calibration attempt 
(Successful). 
NOM.VAL 
Command/Query 
 
Sets or queries the nominal value. The nominal value unit depends 
on the measurement mode. 
Syntax SORT:NOMV<variable><^END^M>or<^J^M>