Gw Instek LCR-816 Digital-Multimeter, DMM, LCR-816 Hoja De Datos

Los códigos de productos
LCR-816
Descargar
Página de 91
  
INTERACE 
117 
 /O_S_REJ 
/O_S_OVER 
The /O_S_REJ or /O_S_OVER signal 
will go low when the secondary 
measurement is over D_Max or 
under D_Min, whilst in C/D , R/Q, 
C/R or L/R mode. The signals will go 
high at time T4     
 /O_EOM 
The End of Measurement signal 
becomes active low when the Bin 
comparison/assignment has 
completed. The signal goes high 
after the next time I_E_TRIG is active 
low.  
Electrical 
Characteristics  
Output Characteristics 
 
 Output 
Voltage 
 
Signal Low 
High 
Max 
current 
/O_BIN1-BIN13 
≤0.5V 
+5V~+24V* 
 
5mA* 
 
/O_S_OVER 
O_S_REJ 
/O_P_OVER 
/O_P_LO 
/O_P_HI 
Control Signals 
/O_INDEX 
/O_EOM 
 
* Pull-up resistors R408~R427 must be replaced to 
output greater than 5V.  
 
   LCR-800 
User 
Manual 
118
 
 
Parameter  Input 
Signals 
 /I_E_TRIG 
Measurement start signal. This 
signal will trigger the LCR-800 to 
start a measurement when the 
signal is pulsed for at least 5us. It 
is triggered by the falling edge of 
the pulse.  
 /I_K_LOCK 
The key lock signal disables the 
panel keys when the signal is low, 
and enables the panel keys when 
the signal is high. 
Electrical 
Characteristics  
Input Characteristics 
 
 
Input Voltage 
Input Current (Low) 
Pull up voltage 
 
Signal Low 
High 5V 12V 
 
/I_E_TRIG 
≤1V 
+5V~15V 
5mA 12mA 
 
/I_K_LOCK 
≤1V 
+5V~15V 
5mA 12mA