6.3.4    CURRent subsystem 
 
The commands in this subsystem are used to configure and control current measurement function and 
are summarized in Table 6-4. 
Table 6-4 
CURRent Subsystem Commands Summary 
Command 
Function Description 
Default   
:CURRent:DC 
 
:NPLCycles <n> 
:NPLCycles? 
:RANGe 
      [:UPPer] <n> 
      [:UPPer]? 
      :AUTO <b> 
      :AUTO? 
    :REFerence <n> 
      :STATe <b> 
      :STATe? 
      :ACQuire 
 
:REFerence? 
Path to configure DC current 
Set integration rate(*power line cycles: 0.1, 1, 10) 
Query line cycle integration rate 
Path to configure measurement range 
Select range (0 to 20) 
Query range 
Enable or disable auto range 
Query auto range 
Specify reference (-20 to 20) 
Enable or disable reference 
Query state of reference 
Use input signal as reference 
Query reference value 
 
 
 
20 
 
ON 
 
OFF 
 
 
:CURRent:AC 
:NPLCycles <n> 
  :NPLCycles? 
  :RANGe 
    [:UPPer] <n> 
    [:UPPer]? 
    :AUTO <b> 
    :AUTO? 
  :REFerence <n> 
    :STATe <b> 
    :STATe? 
    :ACQuire 
  :REFerence?   
Path to configure AC voltage 
Set integration rate (*power line cycles: 0.1, 1, 10) 
Query line cycle integration rate 
Path to set measurement range 
Select range (0 to 20) 
Query range 
Enable or disable auto range 
Query auto range 
specify reference (0 to 20)   
Enable or disable reference 
Query state of reference 
Use input signal as reference 
Query reference value 
 
 
 
20 
 
ON 
 
OFF 
 
 
      
*Fast = 0.1, Medium = 1, Slow = 10 
 
Speed Commands 
 
: NPLCycles <n> 
Command Syntax: 
: CURRent:AC:NPLCycles <n>      
 
    Set NPLC for ACI 
: CURRent:DC:NPLCycles <n>     
 
        Set NPLC for DCI