Bk Precision BK-2831E Digital-Multimeter, DMM, VFD 20000 counts CAT II 300 V BK-2831E Hoja De Datos

Los códigos de productos
BK-2831E
Descargar
Página de 78
57 
 
6.3.5    RESistance subsystem 
The commands in this subsystem are used to configure and control resistance measurement function 
and are summarized in Table 6-5. 
Table 6-5 
RESistance Subsystem Commands Summary 
Command 
Function Description 
Default   
:RESistance 
:NPLCycles <n> 
  :NPLCycles? 
  :RANGe 
    [:UPPer] <n> 
    [:UPPer]? 
    :AUTO <b> 
    :AUTO? 
  :REFerence <n> 
    :STATe <b> 
    :STATe? 
    :ACQuire 
:REFerence?   
Path to configure resistance 
Set integration rate (*power line cycles: 0.1, 1, 10) 
Query line cycle integration rate 
Path to configure measurement range 
Select range (0 to 20e6) 
Query range 
Enable or disable auto range 
Query auto range 
Specify reference (0 to 20e6) 
Enable or disable reference 
Query state of reference 
Use input signal as reference 
Query reference value 
 
 
 
20e6 
 
ON 
 
OFF 
 
 
      
*Fast = 0.1, Medium = 1, Slow = 10 
 
Speed Commands 
 
: NPLCycles <n> 
Command System: 
:RESistance:NPLCycles <n> 
 
 
 
Set NPLC for 
Ω 
 
Command Parameter: 
<n> =    0.1, 1, 10     
 
Set power line cycles per integration 
DEFault          
MINimum 
 
 
10 
MAXimum           0.1 
 
Query: 
:NPLCycles?  
 
 
Query programmed NPLC value 
 
Description: 
The integration period (measurement speed) for the basic measurement functions (except 
Frequency and Period) is set using the  :NPLCycles command.    NPLC  (Number of Power Line 
Cycles) expresses the integration period by basing it on the power line frequency. For example, for a 
PLC of 1, the integration period in seconds would be 1/60 (for 60Hz line power) which is 16.67 ms.