:ACQuire 
Command syntax: 
:RESistance:REFerence:ACQuire   
 
 
Acquire reference for 
Ω 
 
Description: 
When one of these commands is sent, the measurement input signal is acquired and established as 
the reference value. This command is typically used to zero the display. For example, if the 
instrument is displaying a 0.1 
Ω offset, sending this command and enabling reference zeroes the 
display.   
 
This command is functional only if the instrument is on the specified measurement function. 
Sending this command while in any other function causes an error. Also, if the latest reading caused 
an overflow condition or a reading has not been triggered, an error occurs when this command is 
sent. 
 
6.3.6    FREQuency and PERiod subsystem 
The commands in this subsystem are used to configure and control  frequency  as well as  period 
measurement function and are summarized in Table 6-6. 
Table 6-6 
FREQency and PERiod Subsystem Commands Summary 
Command 
Function Description 
Default   
:FREQuency 
:THReshold 
:VOLTage 
    :RANGe <n> 
      :RANGe? 
:REFerence <n> 
    :STATe <b> 
    :STATe? 
    :ACQuire 
:REFerence? 
Path to configure frequency 
Path to select the threshold voltage range: 
 
Select threshold range (0 to 750). 
Query threshold range 
Specify reference (0 to 1.0e6) 
Enable or disable reference 
Query state of reference 
Use input signal as reference 
Query reference value 
 
 
 
20 
 
OFF 
 
 
 
  :PERiod 
:THReshold 
  :VOLTage 
    :RANGe <n> 
        :RANGe? 
:REFerence <n> 
    :STATe <b> 
    :STATe? 
    :ACQuire 
:REFerence?   
Path to configure period 
Path to select the threshold voltage range: 
 
Select threshold range (0 to 750). 
Query threshold range 
Specify reference 
Enable or disable reference 
Query state of reference 
Use input signal as reference 
Query reference value 
 
 
 
20 
 
OFF