Setup Tolerance 
 
1.  Select the primary measurement mode based 
on the type of components to be measured.  
This is done by pressing the 
 (or 
 
for model 878B) button to configure the 
desired measurement mode. 
 
Note:  Be sure to select the correct 
measurement mode, as tolerance mode 
cannot be activated unless the correct mode 
is selected.  For example, if the component is 
a capacitor, be sure to select “C” for 
capacitance.  If not, tolerance mode will not 
be activated following the proceeding steps 
below. 
 
2.  Insert the component to be used as the 
“standard” reference value.  Another words, 
insert a known “good” component that will be 
used for testing against all other components. 
 
(See Figure 6 for illustration) 
 
Note:  The tolerance mode cannot be 
activated unless the meter senses a 
component is connected to either the input 
sockets or terminals.