Bk Precision BK-879B Digital-Multimeter, DMM, LCD 40000 counts BK-879B Hoja De Datos

Los códigos de productos
BK-879B
Descargar
Página de 104
 
 
81 
 
example, if the capacitance of the component is to be 
in the mF range, than selecting 100 Hz or 120 Hz for 
test frequency would give much better results.  The 
results will also be obvious because if the same 
component was tested with 1 kHz or 10 kHz, the 
measured readings may look erroneous on the 
display. 
 
In all cases, it is best to check with the manufacturer’s 
data sheet in order to determine the best test 
frequency to use for measurement. 
 
Inductance 
Typically, a 1 kHz test frequency is used to measure 
inductors that are used in audio and RF circuits.  This 
is because these components operate at higher 
frequencies and require that they be measured at 
higher frequencies such as 1 kHz or 10 kHz.  
However, a 120 Hz test signal is used to measure 
inductors that are used for applications such as filter 
chokes in power supplies, in which are typically 
operated at 60 Hz AC (in U.S.) with 120 Hz filter 
frequencies. 
 
In general, inductors below 2 mH should be measured 
at 1 kHz frequency while inductors above 200 H 
should be measured at 120 Hz.