Capacitance 
When measuring capacitors, it is always most 
desirable if the dissipation factor is low.  Electrolytic 
capacitors inherently have a higher dissipation factor 
due to their normally high internal leakage 
characteristics.  In some cases, if the D (dissipation 
factor) is excessive, measurement accuracy may 
degrade and even read out of specification. 
 
Inductance 
Some inductors are intended to operate at a certain 
DC bias to achieve a certain inductance value.  
However, the 878B and 879B LCR meters cannot 
produce such biasing scheme and external biasing 
should not be attempted because external power 
would be applied to the instrument and cause serious 
damage to the meter.  Therefore, in some cases, 
measured inductance reading may not agree with 
manufacturer’s specification.  It is important to check if 
specification pertains to DC biasing or not. 
 
Resistance 
When measuring resistance of devices, it is important 
to know that there are two types or ways of 
measurement.  One type is DC resistance 
measurement.  Another type is AC resistance 
measurement.  The 879B and 878B uses AC 
resistance measurement method and does not