Operating Manual GLMU 200 MP
page 5 of 8
                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             
Arrow to Temp indicates
temperature display
8 Display 
functions
8.1  Currently measured values
During normal operation the conductivity display value is displayed.
By short press of key ‘SET’ (1) the temperature in [°C] will displayed for 10 seconds.
display of conductivity value
display temperature
8.1.1 Min/Max Value Memory
watch Min values (Lo):
press ‘down‘(2) shortly once
display changes between ‘Lo‘ and Min values
watch Max values (Hi):
press ‘up‘(3) shortly once 
display changes between ‘Hi‘ and Max values
restore current values:
press ‘down‘(2) or ’up’(3) once again
current values are displayed
clear Min-values: 
press ‘down’(2) for 2 seconds
Min values are cleared. The display shows shortly ‘CLr‘.
clear Max-values: 
press ‘up’(3) for 2 seconds
Max values are cleared. The display shows shortly ‘CLr‘.
After 10 seconds the currently measured values will be displayed again.
9  Error and system messages
Display Description
Possible fault cause
Remedy
Err.1
measuring range ex-
ceeded
Wrong signal
Temperature above 140°C not allowed.
Choose right measuring range for con-
ductivity
Err.2
Measuring value below
measuring range
Wrong signal
Temperature below -5°C not allowed.
Choose right measuring range for con-
ductivity
Err.7
System fault
Error in device
Disconnect from supply and reconnect.
If error remains: return to manufacturer
Err.9
Sensor error
Sensor or cable defective Check sensor, cable and connections
Er.11
Calculation not possible
Calculation variable
missing or invalid
Check temperature
8.8.8.8
Segment test
The transducer performs a display test for 2 seconds after power up.
After that it will change to the display of the measuring.