Gw Instek PEL-3021 Electronic Load PEL-3021 Hoja De Datos
Los códigos de productos
PEL-3021
PEL-3000 User Manual
96
8. Choose the P/F-Time (pass/fail time) in
seconds.
The P/F-Time refers to the P/F delay time. This
delay time includes the 0.06 P/F start test time,
as shown in the timing diagram on page 94.
delay time includes the 0.06 P/F start test time,
as shown in the timing diagram on page 94.
P/F-Time
Off, 0.0 ~ 119.9 seconds
9. Set the Short-Time in seconds.
Has the same action as pressing the short key.
See page 61 for details about shorting the load.
See page 61 for details about shorting the load.
Short-Time
Off, 0.1 seconds ~ On-Time
10. Repeat steps 3 to 9 for all the steps in the
program.
A maximum of 16 steps per program can be
created.
created.
Steps that are not configured are set to ―Skip‖
by default.
by default.
11. Press Save[F3] to save the program and all the
steps in the program.
The program will be saved to internal memory.
See the Save/Recall chapter on details on how
to save to Setup memory.
to save to Setup memory.
Recall Default
Pressing Recall Default[F4] will recall the default
settings for each program/step. See page 178
for details.
settings for each program/step. See page 178