Commissioning
Proline Promass 80 PROFIBUS PA
56
TOTAL module 
The measuring device has one totalizer function block. The totalizer values can be cyclically 
transmitted to the PROFIBUS master (Class 1) via the TOTAL module (slot 5). The totalizer value 
is represented in the first four bytes in the form of a floating point number in accordance with the 
IEEE 754 standard. The fifth byte contains standardized status information corresponding to the 
totalizer value. 
Further information on the device status 
Input data
Assignment of the measured variables to the TOTAL module
The TOTAL module can transmit different totalizer variables to the PROFIBUS master (Class 1).
The measured variables are assigned to the Totalizer function block by means of the local display or 
with the aid of an operating program (e.g. FieldCare) in the "CHANNEL" function:
TOTALIZER 
→ TOTALIZER option: select a totalizer → 
CHANNEL: select a measured variable
Possible settings
Factory setting
Example:
You want to cyclically transmit the totalized volume flow as totalizer value 1 to the PROFIBUS 
master (Class 1) by means of the TOTAL module (slot 5):
TOTALIZER 
→ TOTALIZER option: select TOTALIZER 1, then select CHANNEL = VOLUME 
FLOW
Byte 1
Byte 2
Byte 3
Byte 4
Byte 5
Totalizer value (IEEE 754 floating point number)
Status
Totalizer value/measured variable
ID for CHANNEL function
MASS FLOW
277
VOLUME FLOW
273
CORRECTED VOLUME FLOW
398
OFF
0
Module
Totalizer 
function block
Totalizer value/
Measured variable
Unit 
ID for CHANNEL 
function
TOTAL (slot 5)
1
MASS FLOW
kg/s
277