3.6.  Clearing the Stored PPM Log, Alarm and Fault Data 
Up to 200 gas measurements, and 30 alarm and fault events are stored by the monitor.  
To clear  stored data, first display the data to be cleared by using the PPM LOG,  ALARMS  or  FAULTS 
function (page 22). Next, press the Right Arrow Keypad and ENTER buttons at the same time. A single, 
long tone should be heard when the data has been successfully cleared. 
3.7.  Reset to Factory Default Settings 
 
 
IMPORTANT! Performing this function will delete all program parameters, alarms, faults, 
and ppm log files. 
 
With AC power removed from the monitor, press and hold down the Factory Default Switch  (pages 10 
and 11).  Apply AC power and listen for a beep and then a long tone. Release switch. At this time the 
monitor will purge the IR bench and auto set the IR detector voltage to its baseline 0 ppm value. 
After warming up, reprogram the SZ monitor to the desired settings as needed. 
3.8.  The P-CHK Function 
3.8.1.  Overview 
The  P-CHK  function (Pressure Check Function) (page  22)  is useful to a service technician for 
troubleshooting a flow fault problem. The monitor will trigger a flow fault if the pressure drop from ambient 
is less than 0.02 psi during a purge cycle, and 0.06 psi during a measurement cycle.  
3.8.2.  Keypad Functions 
From the Data Display screen, use the Keypad buttons to place the arrow (>) on the display next to the 
P-CHK function. Then press ENTER to display the Pressure screen. 
The  Left Arrow Keypad button toggles the purge valve open and closed. Note that an asterisk (
∗) 
appears when the purge valve is open (purging). 
The Down Arrow Keypad button toggles the pump ON and OFF.  
Pressing the ENTER  button stores the current manifold pressure shown on the left to the ambient 
pressure shown on the right (must be done with the pump OFF). 
3.8.3.  Screen Display 
 
 
Manifold Pressure – Current manifold pressure in psia. 
Stored Ambient Pressure – Stored ambient pressure in psia. 
Pressure Difference  –  The difference between the current manifold pressure and the stored ambient 
pressure. 
Fault Code – Current fault code (page 23). 
Purge Valve Asterisk (
– When viewing the Pressure screen, the purge valve can be opened and 
closed by pressing the Left Arrow Keypad button. An asterisk appears on the display when the purge 
valve is open
 
14.72 * AMB14.81 
 0.09dif  <0000> 
Manifold Pressure    
∗  Stored Ambient Pressure 
Pressure Difference 
 
Fault Code