Figure 3.9 – Calibration Menus 
 
Entries 
Description 
Set Zero 
The CALIBRATION MENU allows for entering of the correct Cal ZERO & Cal SPAN set-
point values needed to calibrate the sensor.  These are entered in the same 
engineering units as input range.  Set Zero & Set Span controls in this menu allow 
pushbutton calibration by moving the pointer to each and pressing the EDIT key.  A 
live reading of the channel’s value allows calibration checks to see if an adjustment 
is needed.  Unintentional calibrations are reset by the Unity Gain menu item. Unity 
Gain  resets zero offset to 0 and span gain to 1.  It is useful for returning the 
calibration to a known starting place.  Sensor aging may be monitored by recording 
zero and span readings at Unity Gain when it is new, and again at later dates when 
degradation may have occurred. 
Set Span 
Cal Zero 
Cal SPAN 
Unity Gain 
Zero Gas 
Span Gas 
 
To check zero calibration, apply the ZERO calibration value to the sensor and observe 
the live reading.  If the zero reading differs from the zero setpoint, a calibration is 
necessary.  To calibrate zero, move the pointer to 
Set Zero and press EDIT.  A warning 
message explains that pressing EDIT again will change the zero calibration and any other 
key will exit.  The procedure for span calibration is identical.  For example, if an LEL 
combustible sensor is to be spanned with 50% LEL span gas, the span set-point must be 
50%.  If 45% LEL is to be used later, the span set-point must be changed to 45% in order 
to match the span calibration gas.  If the reading is only 40% LEL with the 50% gas 
applied, a span calibration is needed.  Move the pointer to the 
Set Span entry and press 
EDIT twice. Unity Gain may be used at anytime to cancel incorrect calibrations and 
start again. 
3.3  System Configuration Menus 
Some items needing configuration are not specific to a channel but affect the entire 
GDA 1600 system.  These are located in the system entry menu shown on the left side of 
Figure 3.10.  System menus are accessed by pointing to the desired item and pressing 
EDIT