Implementing User Calibration
User calibration cannot be conducted in calibration channel CH00.
(CH00 serves as calibration channel for measurement based on KONICA MINOLTA 
SENSING’s calibration standard.) 
When user calibration is performed in the calibration channel for which target color has 
already been set, previous target color is cancelled.
These are commonly used among each color space of L
v
xy, L
v
u’v’, L
v
T∆uv, XYZ, and 
dominant wavelength in one channel.
Operation Procedure
1.
  Press 
ESC
 key when menu or target 
value setting menu appears. 
Measurement screen appears on LCD screen.
 
2. 
 Press 
SHIFT
 key to switch 
to SHIFT mode. 
[SFT] appears on left bottom indicating 
this instrument is to operate in SHIFT 
mode. Or make sure that [SFT] appears on 
left bottom indicating that this instrument 
has been operated in SHIFT mode.
3.
   Press either  
 or 
  key to 
switch calibration channel and 
select one for user calibration.
2
3
1, 10
4, 9
5
5, 7
4, 6
C
S
E
T
T
IN
G
 
CH
01
:
<
DEFAULT
>
  
    Lv  ------
cd/m
2
 
      x  ------
      y  ------       
 
 
<
MODE
>
SINGLE
   [ SFT]      [  2° ]
<
ANGL
>
  1° VIEW
<
SPD
>
  AUTO
<
SYNC
>
NO  SYNC
<
LENS
>
 STANDARD
<
MEM
>
 AUTO NUM
<
MEMORY DATA
>
M000 :(NO DATA)
    Lv  ------
      x  ------
      y  ------
 
CH
01
:
<
DEFAULT
>
  
    Lv  ------
cd/m
2
 
      x  ------
      y  ------       
 
 
<
MODE
>
SINGLE
   [ SFT ]      [  2° ]
<
ANGL
>
  1° VIEW
<
SPD
>
  AUTO
<
SYNC
>
NO  SYNC
<
LENS
>
 STANDARD
<
MEM
>
 AUTO NUM
<
MEMORY DATA
>
M000 :(NO DATA)
    Lv  ------
      x  ------
      y  ------