Frye WinCHAP ver. 2.82 Manual De Usuario

Descargar
Página de 124
96
   Hearing Aid Measurements 
WinCHAP User’s Manual 
7.3.4 Sending curves to your analyzer
You can send any curve displayed in the Response Curves window to your ana-
lyzer using the Send Curves button.
1.  Choose the curves you want to send to your analyzer by having them dis-
played on the window. You can send up to four curves. If you have more 
than four curves displayed, only the first four will be sent.
2.  Click Send Curves. This will open the Send Curves window. See Figure 
7.3.4.
Figure 7.3.4—The Send Curves window
3.  Select Multi-Curve Coupler.
4.  Click Go to send the curves. A window will open, confirming that the 
curves have been sent.
5.  Click 
Back to close the Send Curves window.
7.3.5 Adding saved curves
It can often be useful to compare curves you have previously saved to the 
currently displayed curves. For example, you can compare the response of a 
hearing aid to the response it had at a previous date. Or, you can compare the 
response of the left hearing aid to the saved response of the right hearing aid of 
a binaural set of aids.
1.  Click Add Curves in the Response Curves window. This will open the 
Add Curves window. See Figure 7.3.5. 
2.  Look at the list of displayed hearing aids. If the aid is attached to a client, 
all aids attached to that client will also be displayed. Any aids with saved 
measurements will have a PLUS (+) to the left of the aid listing.
3.  Click the PLUS (+) to the left of the hearing aid listing to display the 
saved tests of the hearing aid. Each test will have a PLUS (+) next to it, 
indicating that there are individual measurements saved to it.