WinCHAP User’s Manual 
Hearing Aid Measurements   
103
•  The ACIC sequence runs the same tests as the ANSI 87 sequence using 
the FONIX CIC coupler and CIC corrections. This test is not, therefore, an 
ANSI S3.22 sequence.
•  The IEC sequence is an automated test used by the international commu-
nity.
•  The JIS sequence is an automated test used primarily in Japan.
 
Figure 7.5—The ANSI 87 automated test sequence window
7.5.1 Getting a test sequence
The most straight forward way to save the automated test sequence results into 
the database is to use the Get Test button:
1.  Set the analyzer up for normal coupler testing.
2.  Run the automated test sequence on your analyzer as you normally 
would.
3.  Click Get Test in the automated test sequence window in WinCHAP. The 
will cause WinCHAP to grab all test data currently displayed on your ana-
lyzer.
4.  Click 
Save in the toolbar to save the results.