WinCHAP User’s Manual 
Audiometric Test Screen   
57
5.  Click Send RIGHT Ear Target Curve to Analyzer or Send LEFT Ear 
Target Curve to Analyzer. This will send the insertion gain target to your 
connected FONIX analyzer.
6.  Click 
Back to exit the target window. 
Figure 5.5.1—Creating a target
5.5.2  Sending an audiogram to the analyzer
If you are using the FONIX 7000 or FP35 analyzer, you can send the WinCHAP 
audiogram to your analyzer. To do this, right click on the audiogram you want 
to send in the Audiometric Curves window and select Send (curve name) to the 
analyzer. This will send the audiogram to the analyzer. You will have to push 
the suitable buttons on the analyzer to actually generate a target. See your ana-
lyzer’s Operator’s Manual for more details on generating a target.
This feature is not available on the 6500-CX or FP40 analyzers.
5.5.3  Starting a hearing aid test
If you have assigned hearing aids to your client (as described in Section 3.2.2.2 
and Section 3.3.2.3), you can go directly to a hearing aid test window from the 
Client & Test Info window.
1.  Click Client & Test Info in the left column of the window.
2.  Look at the bottom half of the Client & Test Info window. You will see a 
list of hearing aids attached to the client.
3.  Double-click on the hearing aid you would like to test. This will open the 
hearing aid test selection window.
4.  Choose the test you would like to perform. The testing window will open.