   The DSL/NAL-NL1 Module 
WinCHAP User’s Manual 
6.2.1.2 To use the audiometry table
1.  Follow steps 1-4 in the previous section.
2.  Select the appropriate column in the Audiometric Information table in the 
lower half of the Audiometry window.
3.  Enter the amplitude data in the boxes corresponding to the appropriate 
frequencies. Use your Enter key to move between fields.
4.  Click 
Save in the toolbar to save the results.
6.2.2 Entering/Predicting Upper Limit values
The AC Upper Limit values are defined as the client’s upper limits of comfort. 
You can choose to have WinCHAP predict these values, based upon the client’s 
age and entered thresholds, or you can enter measured values. WinCHAP uses 
the DSL method to make these predictions.
Choose between Predicted or 
Measured in the Upper Limit box to the right 
of the Assessment Data graph. Measured upper limit values are entered in the 
same manner as threshold values. See Section 6.2.1.
6.2.3 Entering Bone Conduction values
Bone conduction values are entered in the same manner as threshold values. If 
entered, NAL-NL1 uses them in creating measurement targets. DSL does not use 
these values.
6.2.4 Performing or Predicting an RECD measurement
The real-ear to coupler difference (RECD) measurement finds the acoustical dif-
ference between a 2-cc coupler and your patient’s ear. The DSL and NAL-NL1 
formulas take these differences and create realistic coupler target curves. This is 
particularly useful when fitting babies and small children who don’t allow you 
to take an actual real-ear measurement.
To predict the RECD based on the patient’s age, selected Predicted in the 
RECD 
box in the Audiometry window. The predicted RECD (as well as the predicted 
REUR and REDD) values are displayed in column format on the bottom half of 
the window. DSL and NAL use different predictions for these transforms in their 
fitting formulas; the Settings window lets you set which prediction values are 
displayed.
To measure the RECD, select Measured in the RECD box and follow the instruc-
tions in the following sections.
6.2.4.1 To perform the coupler measurement
This section describes how to perform the coupler part of the RECD measure-
ment. The procedure is different for the 6500-CX and FP40 than for the FP35 
and 7000 analyzers.