   The DSL/NAL-NL1 Module 
WinCHAP User’s Manual 
Many of the steps to set up WinCHAP for a DSL verification measurement are 
the same for both coupler and real-ear measurements. These steps are described 
in Section 6.3.1. The actual measurement process is described in Section 6.3.2.
 
Figure 6.3—DSL Measurement window
6.3.1 Setting up
1.  Click Settings and choose the General tab.
2.  Select the Hearing Aid StyleTransducerLoud Speaker Orientation, and 
Compression Threshold.
3.  Choose the DSL Settings tab.
4.  Select the appropriate Circuit Type. Choose between Linear, WDRC 
Fixed, and WDRC Variable.
5.  Select the desired HL to SPL Transform. This affects how the threshold 
values are converted into SPL.
6.  Select whether you want to use a 
Measured or Average RE to 2cc 
Transform in creating the coupler targets. If you choose Measured, you 
will have to perform the RECD measurement in the Audiometry window. 
See Section 6.2.4.
7.  Select whether to use a Predicted or 
Measured Maximum Output. In gen-