Frye WinCHAP ver. 2.82 Manual De Usuario

Descargar
Página de 124
82
   The DSL/NAL-NL1 Module 
WinCHAP User’s Manual 
6.6 Reference
This section describes all of the different settings available in the Settings window.
6.6.1 Understanding General Settings 
•  Hearing Aid Style: Select the hearing aid style affects the natural resonance 
of the ear, and so compensations are made for style when calculating the cou-
pler prescription curves. Choose from BTE, CIC, ITC, ITE, and Body.
•  Transducer: This selection specifies the transducer used in making the 
audiometric measurements. It affects the HL to SPL transform, the targets, 
and some of the other settings. See Table 6.6.1 for a listing of allowed set-
ting combinations. The RECD and HL to SPL settings referred to by the table 
are located in the DSL settings tab. A selection of “Insert Custom Mold” will 
require a measured RECD in order to generate coupler targets in the DSL win-
dow.
•  Loud Speaker Orientation: This selection specifies the position of the loud 
speaker used to obtain audiometric and real-ear sound field measurements. 
We recommend using a 45 degree angle for most repeatable results.
•  Show HL to SPL Transform in: This selection chooses whether to display the 
SPL values in the tables of the Audiometry window as coupler SPL or real-ear 
SPL.
•  Compression Threshold: Choose from 40 dB to 65 dB. Your choice should 
match the compression threshold of the hearing aid.
•  Show Curve Symbols: Turn on or off the display of the numbers and symbols 
identifying the displayed curves.
•  Show Graph Lines: Turn on and off the display of the graph lines in the vari-
ous windows.
•  Share DSL w/NAL Meas: Choose whether or not to make tests taken in the 
DSL window be available in the NAL window.
•  Share NAL w/DSL Meas: Choose whether or not to make tests taken in the 
NAL window be available in the DSL window.
•  Apply Corrections: Choose whether or not to correct the NAL and DSL dB 
SPL targets to take into consideration the FONIX speech weighting. It is rec-
ommended to select Yes.
•  PT Signal Mode: Select the type of signal used with pure-tone measurements. 
The Fast selection does a quick, repeating pure-tone sweep that doesn’t stop 
until you tell it to stop. The Normal selection does one full pure-tone sweep.