Programming and Operating Concepts 
 
Video Recorder – User Manual 
 
77
 
The next step to complete is to initialize the disk on the video recorder as you would have done normally, 
as explained in the «Initializing a zip disk» section, except that this time, it will take just a few seconds. 
 
Note: it is recommended to dedicate  a ZIP disk to storage of data only and store configurations of 
products on a separate disk. 
Data Storage Status
 
The prompt DATA STORAGE STATUS, accessed from the Online DATA STORAGE menu item, 
displays a calculation of how long a particular disk will last based on the configured data storage 
schedule.  Disk capacity is indicated in days, hours, and minutes remaining on the disk. 
After the instrument has been Online and actively performing Data Storage to disk, a warning message 
will appear when the disk reaches the default 90% capacity, or a user-specified capacity.  Once a disk 
has reached its programmed capacity, a DISK FULL message will be displayed. 
Data storage considerations 
In order to guarantee a proper operation of the instrument (e.g. no sample lost), there is a maximum load 
that the video recorder should not overcome. This load can be theorically computed by considering the 
number of data storage trends, live trends and live screens programmed and their associated sample 
rate. 
1 ) One live screen counts for one schedule per second. At least one live screen is present in the 
instrument. A  live screen is a display that require any measurement information to be built. 
2 ) Each ENABLED Data Storage schedule has a user defined sample rate. 
3 ) Each ENABLED Live Trend has the following sample rates :  
 
Screen Size 
Schedule Sample Rate 
 
5 Min Screen 
1 Second 
15 Minute Screen 
3 Seconds 
30 Minute Screen 
6 Seconds 
1 Hour 
12 Seconds 
2 Hours 
24 Seconds 
4 Hours 
47 Seconds 
8 Hours 
93 Seconds (1.55 Minutes) 
24 Hours 
279 Seconds (4.65 Minutes) 
7 Days 
1951 Seconds (32.5 Minutes) 
31 Days 
8640 Seconds (2.4 Hours) 
 
Therefore, to guarantee a proper operation of the instrument, the following inequation should allways be 
true : 
Σ
P ro g ra m m e d   tre n d
(liv e   o r  d a ta   s to ra g e )
A s s o c ia te d  s a m p le  ra te
N u m b e r  o f
liv e   s c re e n s
+
<   6