2.4
Configuration
2.4.1
AE Title/ Presentation Address Mapping
2.4.1.1
Local AE Titles
"Workstation" uses the AE Titles and TCP/IP Ports configured by means of the  "Workstation Service & 
Configuration Tool". The Field Service Engineer can configure the TCP Port via the "Workstation Service & 
Configuration Tool". No Default AE Titles are provided. The AE Titles must be configured during installation. 
The local AE Title used by each individual application can be configured independently of the AE Title used 
by other local applications.
Table 2.4-1: AE Title Configuration Table
Application Entity
Default AE Title
Default TCP/IP Port
Storage
No Default
104
Hardcopy
No Default
Not Applicable
RIS
No Default
Not Applicable
MPPS
No Default
Not Applicable
2.4.1.2
Remote AE Title/ Presentation Address Mapping 
The AE title, host names and port numbers of remote applications are configured using the "Workstation 
Service & Configuration Tool". 
2.4.1.2.1
Storage
The  "Workstation Service & Configuration Tool" must be used to set the AE Titles, port-numbers, host-
names and capabilities for the remote Storage SCPs. Associations will only be accepted from known AE 
Titles while associations from unknown AE Titles will be rejected (an AE Title is known if it can be selected 
within the "Workstation Service & Configuration Tool"). Multiple remote Storage SCPs can be defined. Any 
Storage SCP can be configured to be an “Archive” device causing storage commitment to be requested for 
images or presentation states transmitted to the device.
Storage Commit can be enabled or disabled.
The Archive Connection can be tested by means of a DICOM Ping (C-echo) test to the archive.
2.4.1.2.1.1
Archive options
When sending images to an archive, this can be done by means of the following SOP classes: CR, DX and 
MG for presentation or DX and MG for processing. The latter is a RAW image, without any annotations or 
measurements burned in.
Image pixels can be sent in the requested output type (configurable) depending on the used SOP class and 
whether GSPS is supported or not
CR image, WITHOUT GSPS:
-
8 bit OD
-
8 bit OD Gamma correction
-
8 bit P-Value
 
 
6
anonymous archiving (i.e. without patient demographics) is not supported