Philips MP20/30 Manuel D’Utilisation

Page de 414
53
4
4
Patient Alarms and INOPs
This chapter lists patient alarms and technical alarms (INOPs) alphabetically, irrespective of their 
priority. INOPs start on page 59.
For information on alarms and INOPs from the gas modules refer to your Gas Module Instructions for 
Use.
Patient Alarm Messages
The measurement labels and abbreviations for pressure, temperature, SpO
2
, and anesthetic agent 
alarms are explained in the individual chapters. 
Some alarms may be shown at the Information Center in shortened form, when transferred through 
IntelliVue Instrument Telemetry (USA only). These shortened alarm texts are included in the list and 
identified with the note “at Information Center”.
Note that yellow arrhythmia alarms (“short yellow alarms”) may be shown with one or with two stars, 
depending on your monitor configuration and the Information Center revision you are using.
Refer to your Gas Module Instructions for Use for patient alarms and INOPs from the gas modules.
Alarm Message
From
Condition
Indication
***APNEA 
or
***APNEA xxx sec
CO
2
, Resp, 
AGM
Respiration has stopped for longer than 
the preset apnea time. “xxx” denotes the 
Apnea duration.
numeric flashes, red alarm 
lamp, alarm tone. 
***ASYSTOLE
ECG
No QRS detected for a period greater 
than the asystole threshold (in the 
absence of Vfib or chaotic ECG).
numeric flashes, red alarm 
lamp, alarm tone.
**awRR HIGH
CO
2
, Resp, 
AGM
The airway respiration rate has exceeded 
the high alarm limit.
numeric flashes and high limit 
is highlighted, yellow alarm 
lamp, alarm tone.
**awRR LOW
CO
2
, Resp, 
AGM
The airway respiration rate has dropped 
below the low alarm limit.
numeric flashes and low limit 
is highlighted, yellow alarm 
lamp, alarm tone.
**BIS HIGH
BIS
The Bispectral Index value has exceeded 
the high alarm limit.
numeric flashes and high limit 
is highlighted, yellow alarm 
lamp, alarm tone.
**BIS LOW
BIS
The Bispectral Index value has dropped 
below the low alarm limit.
numeric flashes and low limit 
is highlighted, yellow alarm 
lamp, alarm tone.