Portable Oxygen Monitor  
Appendix 
 
 
 Teledyne 
Analytical 
Instruments 
43 
Appendix 
A.1 Specifications 
 Range:   0-100% oxygen 
 Accuracy:  +2% of full scale (at constant temperature 
and pressure) 
 Response 
Time:   90% in less than 8 seconds at 25 °C 
 Battery 
Life:  Approximately 2000 hr. continuous use in 
a non-alarm condition 
 System 
Power: 
3 AA alkaline batteries. 
 Sensor 
Type: 
Class R17MED 
 
Expected Life:  36 months in air. (10 months when 
continuously exposed to 100% oxygen) 
 Dimensions:  2.5" W × 1.25" D × 4.5" H  (66 mm x 33-
mm × 111.5 mm) 
 Sensor 
Cable:   Retracted: 2 ft / Extended: 10 ft.  
                    Storage Temp.  10-30 °C (continuous), 5-50 °C  
(Intermittent)   
 
Operating Temp:   0-40 °C 
  
 Alarm 
ranges:  HI alarm: 19 to 100 Percent 
                  
   LO alarm: 18 to 99 % 
                 Alarm accuracy:   Alarm thresholds are digitally stored and 
alarm affectivity is digitally compared 
against the value calculated by the 
microprocessor. Hence, the accuracy of the 
alarms is the same as the accuracy of the unit 
as specified above.