Emerson 20002158 Manuel D’Utilisation

Page de 36
Core processor terminals
Figure 3-3:   
Connect the wires to the transmitter 
terminals
(see the transmitter manual)
Reinstall and tighten the core processor cover
Core processor 
type
From Step 1 or 2
Connect the wires to the core processor terminals:
Red wire
>
Terminal 1 (Power supply +)
Black wire
>
Terminal 2 (Power supply –)
White wire
>
Terminal 3 (RS-485/A)
Green wire
>
Terminal 4 (RS-485/B)
Connect the wires to the core processor terminals:
Red wire
>
Terminal 1 (Power supply +)
Black wire
>
Terminal 2 (Power supply –)
White wire
>
Terminal 3 (RS-485/A)
Green wire
>
Terminal 4 (RS-485/B)
Enhanced 
core processor
Standard 
core processor
1. Reinstall the core processor cover.
2. Torque cover screws to:
10–13 in-lbs (1,13–1,47 N-m) for aluminum housing
Minimum 19 in-lbs (2,1 N-m) for stainless steel 
housing
3.3
Connect 9-wire cable
1.
Prepare and install the cable according to the instructions in the Micro Motion 9-
Wire Flowmeter Cable Preparation and Installation Guide
.
2.
Insert the stripped ends of the individual wires into the terminal blocks. Ensure that
no bare wires remain exposed.
Wiring
26
Micro Motion ELITE