5.4.3 Standby 
 
If this command is enabled, part of the CCD circuitry shuts down (Standby), in order 
to optimize low power consumption when not reading. When the trigger is pressed 
this circuitry powers up. This mode causes a minor delay of about 100 ms before the 
reader is ready. 
 
For M-series readers, standby is always enabled and is not configurable. To exit 
Standby press the trigger. 
 
 
5.5 READING 
PARAMETERS 
 
5.5.1 Operating 
Mode 
 
This group of parameters allows setting different reading modes for either Hand-Held 
operation or Stand operation: 
 
• 
SoftwareTrigger: the reader is set in FLASH MODE. Code reading takes place 
during the flash on time; 
 
• 
Hardware Trigger: the code reading is started with a trigger press (ON); 
 
• 
Hardware Trigger Ready: the illuminator is switched ON when the reader sees a 
code. It functions as a pointer aiming at the code to be read. The reading phase 
starts only when the trigger is pressed. In this mode the reader is automatically 
set to trigger active level and one read per cycle parameters. (Only for Hand-
held operation). 
 
• 
Automatic: The illuminator is switched ON when the reader sees a code. The 
reading phase starts automatically. 
 
• 
Always ON: the illuminator is always ON and the reader always ready for code 
reading. This value is not available for M-series readers. 
 
 
5.5.2 
Hardware Trigger Mode 
 
This mode determines how the reading phase is controlled when the hardware 
trigger operating mode is selected: 
• 
trigger active level: the reader goes ON when the trigger is pressed and goes 
OFF when it is released 
• 
trigger active pulse: the reader goes ON at the first trigger press and goes OFF 
only at a second press