Siemens Software version syngo CT 2005C Manuel D’Utilisation

Page de 321
Thorax
109
 
Hints
• You could repeat the same protocol by simply click-
ing the chronicle with the right mouse button for 
“repeat“, e.g., when both non-contrast and contrast 
studies are required.
• For lung cancer evaluation, this protocol can be com-
bined with the protocol “Neck Routine”.
• Low dose lung images are usually evaluated using 
lung window setting. Soft tissue/bone window set-
tings may be used to visualize the presence of calci-
fications in the nodules.
• It is essential to use the same protocol for follow-up 
studies to check for progression.
LungLowDose
kV
130
Effective mAs
30
Rotation time
1.0 sec
Slice collimation
5.0 mm
Slice width
8.0 mm
Feed/Rotation
10.0 mm
Pitch Factor
1.0
Increment
8.0 mm
Kernel
B70s
CTDI
Vol
3.23 mGy
Effective dose
Male: 1.69 mSv
 
Female: 2.08 mSv
Contrast medium IV injection
Start delay
30 sec.
Flow rate
2.5 ml/sec.
Total amount
50 – 70 ml
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book  Page 109  Friday, April 8, 2005  9:55 AM