Starting therapy using oxygen
1
Fit the oxygen connector to the oxygen inlet of the device.
2
Attach the other end of the oxygen supply tubing to the oxygen supply.
3   
Press 
 to start treatment.
4
Turn on oxygen.
Stopping therapy using oxygen
1
Turn off oxygen.
2
Press  
to 
stop treatment.
Using the FiO
2
 monitoring sensor
Use of the FiO
2
 monitoring sensor may be recommended by your clinician.
CAUTION
Do not use the FiO
2
 monitoring sensor with the H4i humidifier.
Preparing to use a new sensor
1
Before use, leave the FiO
2
 monitoring sensor open to the air for 15 minutes.
2
Attach a new FiO
2
 monitoring sensor (as shown below).
3
Perform the sensor calibration (see “Setup menu” on page 16).
Note: The FiO
2
 monitoring sensor must be replaced every 12 months.
1
2
Oxygen 
supply
Press spring to release
connection after use
1
2
1
2
3
4
5