4.4 
mV Calibration (Offset) (Only for pH 110) 
With pH 110 meter, you can perform a calibration or an 
offset of up to ±150 mV.  Note: For Oxidation reduction 
Potential (ORP) or Redox measurements, an ORP 
electrode must be used. 
1.  While in the measurement function, press 
MODE to enter the mV mode. The mV indicator 
appears in the upper right hand corner.  See 
figure at right.  
2.  Press the CAL/MEAS key. The CAL indicator 
appears above the primary display. The primary 
display shows the relative mV reading and the 
secondary display shows the absolute mV value. 
• 
If you have never calibrated relative mV or if the 
meter has been reset, the value shown in the 
primary display is the same as the absolute mV 
value.  
• 
“R.” annunciator will appear once mV calibration 
is performed, an indication of mV offset. 
3.  Press the MI/▲ or MR/▼keys to enter the 
relative mV value that matches your desired 
reading.  
4.  Press the HOLD/ENTER key to confirm the 
reading and to return to the measurement mode. 
The primary display now shows the relative mV 
reading.  
 
 
 
NOTES: 
1.  To view the mV offset value, see section 9.4.2 - View mV 
Offset on page 40.  
2.  To reset all calibration and offset values in memory to the 
factory default settings, see section 9.6.1 - Calibration Reset 
page 42.  
 