Hanna Instruments hi 983322-0 Manuel D’Utilisation

Page de 2
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
POWER CONNECTION
HI 983322-0 and HI 98332200
HI 983322-0 and HI 98332200
HI 983322-0 and HI 98332200
HI 983322-0 and HI 98332200
HI 983322-0 and HI 98332200
Connect a 2-wire power cable to the terminal strip while
paying attention to the correct positive and negative con-
tacts (12 VDC).
HI 983322-1
HI 983322-1
HI 983322-1
HI 983322-1
HI 983322-1
Connect a 3-wire power cable to the terminal strip while
paying attention to the correct earth, neutral and line
contacts (115/230 VAC).
Notes:
Notes:
Notes:
Notes:
Notes:
• All external cables connected to the rear panel should
   end with wire lugs.
• It is recommended to cover the unused terminals with
   insulating tape.
• At start up the meter needs a few seconds to stabilize.
   Wait until a stable reading is displayed.
ALARM CONTACT
Use this contact (maximum 5A, 240VAC, 30VDC) for connec-
tion to an alarm or dosing system. The unit acts as a switch
to control an external device.
OPERATING THE METER
All operations can be controlled via front panel keys and
trimmers.
“SET” and “MEAS” LEDs light up to indicate which is the
operating function.
Make sure that the meter is calibrated and the Setpoint is
properly selected  before performing any measurement.
Attach the probe to the meter. Install the probe in the
fittings or immerse it in the solution to be monitored, while
making sure that metal pins are completely submerged.
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
HI 7634-00 EC/TDS probe
HI 7061M Electrode cleaning solution, 230mL bottle
HI 710005 12 VDC power adapter, US plug
HI 710006 12 VDC power adapter, European plug
HI 710012 12 VDC power adapter, Australian plug
HI 710013 12 VDC power adapter, Southern Africa plug
HI 710014 12 VDC power adapter, UK plug
HI 731326 Calibration Screwdriver (20 pcs)
HI 740146 Mounting brackets
Front view of the
panel-mounted unit
Side view of the
panel-mounted unit.
Supplied adjustable lo-
cation brackets allow the
controller to slide into the
cutout and will hold the
unit securely in place. 95
mm (3.74") is the mini-
mum  amount of space
required to install the
controller.
Panel-
mounted unit
assembling
view
73mm
2.87"
79mm
3.11"
42mm
1.65"
49mm
1.93"
95mm MIN
3.74"
76mm
3"
0.25/4mm
0.01/0.160"
79mm
3.11"
ADJUSTABLE
LOCATION
BRACKET
HI983322 LAYOUT
HI983322 LAYOUT
HI983322 LAYOUT
HI983322 LAYOUT
HI983322 LAYOUT
HI 740146
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FRONT PANEL
FRONT PANEL
FRONT PANEL
FRONT PANEL
FRONT PANEL
Keypad
SET
SET
SET
SET
SET
To display
setpoint
MEAS
MEAS
MEAS
MEAS
MEAS
To display
measurement
Trimmers
CAL
CAL
CAL
CAL
CAL
For calibration
SET
SET
SET
SET
SET
To adjust
setpoint
LEDs
SET
SET
SET
SET
SET
ON when LCD displays the set value
MEAS
MEAS
MEAS
MEAS
MEAS
ON when LCD displays the measured value
ALARM
ALARM
ALARM
ALARM
ALARM
ON when alarm contact is activated
REAR  PANEL
REAR  PANEL
REAR  PANEL
REAR  PANEL
REAR  PANEL
1. Power Supply:
  HI 983322-0              HI 983322-1
  HI 983322-0              HI 983322-1
  HI 983322-0              HI 983322-1
  HI 983322-0              HI 983322-1
  HI 983322-0              HI 983322-1
   HI 98332200
   HI 98332200
   HI 98332200
   HI 98332200
   HI 98332200
          12 VDC
          12 VDC
          12 VDC
          12 VDC
          12 VDC                 115/230 VAC
                 115/230 VAC
                 115/230 VAC
                 115/230 VAC
                 115/230 VAC
L1:
L1:
L1:
L1:
L1:
Positive                    230 VAC
L2:
L2:
L2:
L2:
L2:
Negative                  115 VAC
L3:
L3:
L3:
L3:
L3:
 ------                      Neutral
L4:
L4:
L4:
L4:
L4:
 ------                      Protection Earth
Note
Note
Note
Note
Note: : : : : the power input is internally protected by a 400 mA fuse
2. Alarm  Contact: A1, A2.
This contact acts as a switch for the power. The contact
has to be protected outside by the user.
3. Probe Connection: follow the above diagram and connect the
     colored wires of probe cable as indicated. It is recommended
    to connect the shield (P5) to avoid any interference.
Press the “MEAS” key.
The LCD will show the conductivity value of the solution in
µS/cm unit. Any initial variation on readings may be due
to temperature compensation.
The “ALARM” LED will light up when the alarm contact is
losed, to indicate an EC value higher than selected setpoint.
CALIBRATION
This meter is factory calibrated.
If re-calibration becomes necessary, a calibration solution
has to be prepared with the help of HI 9033 EC-meter (or
similar). Recommended conductivity value of calibration solu-
tion should be around 13 µS/cm.
Make sure the meter is in the measurement mode (the
“MEAS” LED lights on).
Immerse the probe in the prepared solution.
Shake briefly and wait for reading to stabilize.
Using a small screwdriver adjust the calibration trimmer
until the meter displays the required EC value on the LCD.
SETPOINT
Press the “SET” key. The display will show the default or
previously adjusted value for the setpoint.
Using a small screwdriver, adjust the “SET” trimmer until
the required limit value is displayed.
PROBE MAINTENANCE
To improve probe performance and prolong its life, it is
recommended to clean it regularly.
• Immerse the tip of the probe in HI 7061 Cleaning
Solution at least for one hour.
• If a more thorough cleaning is required, brush the
metal pins with very fine sandpaper.
• After cleaning, rinse the probe with tap water and
recalibrate the meter.
•  When not in use, clean the probe before storing it.