FUNCTIONAL DESCRIPTION
FRONT PANEL
Keypad
S E T
To display
setpoint
MEAS
To display
measurement
Trimmers
CAL
For pH Offset
Calibration
S E T
To adjust the
Setpoint
LEDs
SET
ON when LCD
displays  the set  value
MEAS
ON when the LCD displays the measurement
value
DOSING
ON when dosing contact is activated
REAR PANEL
1. Power Supply:
  HI 981411-0        HI 981411-1
        12VDC              115/230 VAC
L1:
Positive
230 VAC
L2:
Negative
115 VAC
L3:
------
Neutral
L4:
------
Protection Earth
Note: the power input is internally protected by a 400 mA fuse.
2 . Dosing Selection Contacts:
Leave S1 and S2 open to select ACID dosage;
Short S1 and S2 to select ALKALINE dosage.
3 . The dosing (alarm) contact acts as a switch. The
contact has to be protected outside by the user.
HI 981411
pH  CONTROLLER
SET
MEAS
SET
DOSING
CAL
pH
0.0 ...14.0 pH
OPERATIONAL GUIDE
POWER CONNECTION
HI 981411-0
Connect a 2-wire power cable to the terminal strip while
paying attention to the correct positive and negative con-
tacts (12VDC).
HI 981411-1
Connect a 3-wire power cable to the terminal strip while
paying attention to the correct earth, neutral and line
contacts (115/230 VAC).
Notes:
• All external cables connected to the rear panel should
   end with wire lugs.
• It is recommended to cover the unused terminals with
   insulating tape.
• At start up the meter needs a few seconds to stabilize.
   Wait until a stable reading is displayed.
DOSING CONTACT
This contact (maximum 5A, 240VAC, 30VDC) is used for
connection to an alarm or dosing system. The relay contact
acts as a power switch for the controlling device.
OPERATING THE METER
All settings of the parameters can be controlled via front
panel keys and trimmers.  “SET” and “MEAS” LEDs light up
to indicate which is the operating function.
Make sure that the meter is calibrated and that the
Setpoint is properly adjusted before proceeding.
Attach the pH-electrode to the meter with a BNC connector.
Install the electrode in the fittings or immerse it in the
solution to be monitored. Press the "MEAS" key.
The LCD will show the pH value. Any initial variation on
readings may be due to temperature compensation.
The “DOSING” LED will light up when the dosing contact is
closed.
CALIBRATION
Make sure the meter is in the measurement mode (the
“MEAS” LED lights on). Immerse the electrode in HI 7007
(pH7.01) buffer solution (or any buffer solution close to the
operating pH value). Shake briefly and wait for reading to
stabilize. Adjust the calibration trimmer until “7.0” is
displayed on the LCD.
SETPOINT
Press the "SET" key. The display will show the default or
previously adjusted value. Using a small screwdriver adjust
Front view of the
panel-mounted unit
Side view of the
panel-mounted unit
Supplied adjustable lo-
cation brackets allow
the controller to slide
into the cutout and will
hold the unit securely
in place. 95mm (3.74")
is the minimum amount
of space required to
install the controller.
Panel-mounted unit assembling view
73mm
2.87"
79mm
3.11"
42mm
1.65"
49mm
1.93"
95mm MIN
3.74"
76mm
3"
0.25/4mm
0.01/0.160"
79mm
3.11"
ADJUSTABLE
LOCATION
BRACKET
HI981411 LAYOUT
HI 740146
ACCESSORIES
pH CALIBRATION SOLUTIONS
HI 7007M
pH 7.01 Buffer Solution, 230mL bottle
HI 7007L
pH 7.01 Buffer Solution, 460mL bottle
ELECTRODE STORAGE SOLUTION
HI  70300M
Storage Solution, 230mL bottle
ELECTRODE CLEANING SOLUTIONS
HI  7061M
General Cleaning Solution, 230mL bottle
HI  7061L
Oil & Fat  Cleaning Solution, 230mL bottle
pH ELECTRODES
HI1002/5
Heavy-duty BNC-type double junction, plastic-
body with external thread and 5m (16.5’) cable
HI1090T
Screw type, double junction, glass-body with
external thread
HI1110S
Screw type, single junction, glass-body
HI1210T
Screw type, double junction, plastic-body with
external thread
HI2114P/2 BNC-type, double junction plastic-body  with
2m (6.6’) cable
HI1210B/5 BNC-type cable, double  junction, plastic-body,
with 5 m (16.5') cable
HI2910B/5 BNC-type,  double  junction, plastic-body, built-
in amplifier, with  5 m (16.5') cable
EXTENSION CABLES FOR SCREW-TYPE ELECTRODES
TO BNC CONNECTOR
HI7855/5
Extension cable 5m (16.5') long
HI7855/10
Extension cable 10m (33') long
OTHER ACCESSORIES
BL PUMPS Dosing Pumps with flow rate from 1.5 to 20 LPH
HI 6050
Submersible  electrode holder, 60cm (24”)
HI 6051
Submersible electrode holder, 110cm (43”)
HI 6054B
Electrode Holders for in-line applications
HI 6054T
Electrode Holders for in-line applications
HI 710005
12 VDC power adapter, US plug
HI 710006
12 VDC power adapter, European plug
HI 710012
12 VDC power adapter, Australian plug
HI 710013
12 VDC power adapter,Southern Africa plug
HI 710014     12 VDC power adapter, UK plug
HI 731326
Calibration Screwdriver (20 pcs)
HI 7871
Level Controller (min and max)
HI 7873
Level Controller (min, max and overflow)