Hanna Instruments hi 983305 Manuel D’Utilisation

Page de 2
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range
0 to 20 µS/cm (EC)
Accuracy (@ 25°C/77°F)
±2% f.s.
Typical EMC Deviation
±2% f.s.
Temperature
Automatic  from 5 to 50°C (41 to 122°F)
Compensation
with ß = 2.4%
Setpoint
1 to 5 µS/cm
Hysteresis
1 µS/cm over setpoint
Alarm
LED blinks and relay is activated when
EC value is 1 µS/cm higher than setpoint
Relay
SPST NC contact output, 0.5A 30VDC/48VAC max,
resistive load, 100,000 strokes
Probe
HI 7631/2 EC probe (included)
Casing
IP54
Power supply
External 12 VDC (included)
Dimensions
86 x 94 x 33 mm (3.4 x 3.7 x 1.3”)
Weight
150 g (5.3 oz.)
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
• Turn the meter on by connecting the 12 VDC power
adapter to the meter and to the mains.
• Immerse the EC probe in the solution,
making sure that metal pins are com-
pletely submerged.
Note: In order not to affect the accuracy of
measurements, it is important that the
probe body does not touch nor stand
close to the side walls of the vessel. The tip
may lay on the bottom of the beaker.
• Adjust the setpoint knob to the desired
value in the range 1 to 5 µS/cm.
• When the EC reading becomes 1
µS/cm higher than setpoint, the
red alarm LED will blink and output relay is activated.
Note: During alarm condition the relay is de-energized (the
alarm contacts are open). When not in alarm condition,
the relay is energized (the alarm contacts close). If the
power supply is interrupted, the relay is de-energized as
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
HI 7631/2 EC “flow thru” probe with ½”NPT thread, ATC
and 2 m (6.6’) cable
HI 7061L Electrode cleaning solution, 460 mL bottle
HI 710005 12 VDC power adapter, US plug
HI 710006 12 VDC power adapter, European plug
HI 710012 12 VDC power adapter, Australian plug
HI 710013 12 VDC power adapter, Southern Africa plug
HI 710014 12 VDC power adapter, UK plug
if in alarm condition to alert the operator.
PROBE MAINTENANCE
This meter does not need probe cleaning when used in pure
water applications. If the probe needs to be cleaned, the
following procedure should be performed.
• Immerse the tip of the electrode in HI 7061 Cleaning
Solution for one hour.
• If a more thorough cleaning is required, brush the metal
pins with very fine sandpaper.
• After cleaning or when not in use, rinse the probe with
tap water.
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure
that they are entirely suitable for the envi-
ronment in which they are used. Operation
of these instruments in residential areas
could cause unacceptable interferences to
radio and TV equipment. The metal band at
the end of the probe is sensitive to electro-
static discharges. Avoid touching this metal
band at all times.
During operation, ESD wrist straps should be
worn to avoid possible damage to the probe by
electrostatic discharges. Any variation intro-
duced by the user to the supplied equipment
may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use these in-
struments when voltages at the measurement
surface exceed 24 VAC or 60 VDC. Use plastic
beakers to minimize any EMC interferences. To
avoid damages or burns, do not perform any measurement in microwave ovens.
WARRANTY
WARRANTY
WARRANTY
WARRANTY
WARRANTY
HI983305 is warranted for two years against defects in workmanship and materials when
used for its intended purpose and maintained according to instructions. The probe is war-
ranted for a period of one year. This warranty is limited to repair or replacement free of
charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under
warranty, report the model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authoriza-
tion Number from the Customer Service department and then send it with shipment costs prepaid.
When shipping any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the
date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent
of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
I
S983305R1  06/98
PRINTED IN PORTUGAL