eutech-instruments fluoride epoxy Manuel D’Utilisation

Page de 22
Instruction Manual 
 
Fluoride Electrode
 
 
21
                         
to complexing agents                        
Complexation
 
                        
air bubble on membrane          
 
remove bubble by re-dipping 
                         
 
 
 
 
 
probe 
 
"Incorrect Answer"    incorrect scaling         
 
plot millivolts on the linear axis. 
(but calibration       
of semi-log paper        
 
On the log axis, be sure that 
curve is good)                                
 
 
 
concentration numbers within 
                                          
 
 
 
 
each decade are increasing with 
                                           
 
 
 
 
increasing concentration 
 
                        
incorrect sign          
 
 
be sure to note sign of  
                                               
 
 
 
millivolt number correctly 
                                             
                         
incorrect standards       
 
prepare fresh standards 
 
                         
wrong units used          
 
apply correct conversion factor: 
                                                
 
 
 
10
-3
M = 19 ppm as F
-1
 
 
                         
complexing agents       
 
check section entitled 
                         
in sample         
 
 
Complexation