eutech-instruments field phorpconductivity monitors Manuel D’Utilisation

Page de 19
 
 
12
11 SPECIFICATIONS 
Conductivity 
SPECIFICATIONS pH 
ORP 
FM 
CON1 
FM 
CON2 
 
FM 
CON3 
Range 
0.0 to 14.0 pH  
0 to 995 mV  
0 to 10.0 
mS 
0.0 to 
990 
µS 
0.0 to 
150 
µS 
 
Accuracy 
± 0.1 pH  
± 5 mV  
±0.2 
mS/cm  
±20 
µS/cm 
±2 
µS/cm  
 
Resolution 
0.1 pH  
5 mV 
0.1 
mS/cm 
10 
µS/cm 
µS/cm 
Calibration 
Up to 3 points;  
pH 4.0, 7.0 & 10.0 
Single- point, 
Offset ± 100 mV  
Single- point, 
± 30% full scale 
Automatic Buffer Recognition Yes 
 
 
Automatic Temperature 
Compensation  
 
 Yes 
Operating Temperature  
0 to 50 °C (0 to 120 °F) 
Inputs BNC 
BNC 
Permanently 
wired 
Display 
3-digit LED (Red) 
Environmental Requirements 
Operating: 0 to 50 °C (0 to 120 °F) 
Storage: - 10 to 60 °C (14 to 140°F) 
Humidity Limits: 10 to 95% RH (non condensing) 
Power Requirements 
AC / DC Adapter 9 V, 200 mA 
Dimension / Weight 
Meter: 19 x 10 x 6 cm; 80 g 
Boxed: 24 x 23 x 7 cm; 550 g