Summit DL1KIT Guide De Démarrage
Preparing the ACCUCOLD™ Temperature Logger for Use
Before you can use the Temperature Logger, you will have to insert the CR2032 3-volt coin cell,
which is supplied separately with your appliance. To do this, use a Phillips head screwdriver to
loosen and remove one screw from the small plate on the top of the Temperature Logger. (See
illustration below.) Slide the plate out of the way and insert the coin cell, face down. Slide the
plate back and screw it into place.
The Temperature Logger will be automatically energized when you plug your appliance into a
115V AC outlet.
Before use, the ACCUCOLD™ flash drive must be activated. See the included instruction sheet
to set up and activate the flash drive using your computer. When you are ready to begin
recording data, insert the ACCUCOLD™ flash drive into its socket at the front of the
Temperature Logger.
which is supplied separately with your appliance. To do this, use a Phillips head screwdriver to
loosen and remove one screw from the small plate on the top of the Temperature Logger. (See
illustration below.) Slide the plate out of the way and insert the coin cell, face down. Slide the
plate back and screw it into place.
The Temperature Logger will be automatically energized when you plug your appliance into a
115V AC outlet.
Before use, the ACCUCOLD™ flash drive must be activated. See the included instruction sheet
to set up and activate the flash drive using your computer. When you are ready to begin
recording data, insert the ACCUCOLD™ flash drive into its socket at the front of the
Temperature Logger.
Troubleshooting
If the blue indicator light on the front of the Temperature Logger is on steady for a protracted
period of time, the Logger is no longer recording. This may be due to several factors:
period of time, the Logger is no longer recording. This may be due to several factors:
1.
The flash drive is not fully reinserted after a drive removal.
Make sure the drive is inserted all the way. It will take 20 seconds for the device to start
recording again, indicated by the light flashing every second.
Make sure the drive is inserted all the way. It will take 20 seconds for the device to start
recording again, indicated by the light flashing every second.
2.
The flash drive is full.
Transfer or erase the stored data as described on the Instruction Sheet.
Transfer or erase the stored data as described on the Instruction Sheet.
3.
The flash drive is not in the correct format.
Format the flash drive to the FAT file system, use the ACCUCOLD™ application to set the
time and assign the time file to the formatted drive.
Format the flash drive to the FAT file system, use the ACCUCOLD™ application to set the
time and assign the time file to the formatted drive.
4.
The flash drive has encountered a problem in recording the data.
Remove the flash drive and reset the Temperature Logger by unplugging the AC adapter
from the back of the Logger, then reconnecting. Re-insert the flash drive. The indicator light
should begin flashing after 20 seconds.
Remove the flash drive and reset the Temperature Logger by unplugging the AC adapter
from the back of the Logger, then reconnecting. Re-insert the flash drive. The indicator light
should begin flashing after 20 seconds.