Graham Field 760000-INS-LAB-RevA08 Manuel D’Utilisation
760000-INS-LAB-RevA08
13
general maIntenanCe
1. Check main power cord for abrasion or excessive wear.
2. Check mattress cover for signs of wear or damage. En-
2. Check mattress cover for signs of wear or damage. En-
sure mattress cover and tubes are connected correctly.
3. Check airflow from the air hose connector. The airflow
should alternate between each connector every half-
cycle time if it’s in alternating mode.
cycle time if it’s in alternating mode.
4. Check the air hoses for any kink or break. For replace-
ment, please contact your GF equipment provider.
Fuse replacement
tool needed: small screwdriver
1. If you suspect a blown fuse, discon-
nect the plug from the wall outlet
immediately.
immediately.
2. Use a small screwdriver to remove
the cover of the fuse holder (fuse locations are shown
in picture at right).
in picture at right).
3. Insert a new fuse of the correct rating (T1A/250V, VDE
approved).
4. Replace the fuse holder cover. Ensure that fuse holder
cover is securely installed.
Air filter replacement
tool needed: small screwdriver
1. Use a small screwdriver to remove
the air filter plate located at rear
of pump (filter location shown in
picture at right).
of pump (filter location shown in
picture at right).
2. Remove the filter. The filter is reusable, if not torn, and
can be washed gently with a mild detergent and water.
Dry the filter before use.
Dry the filter before use.
3. Replace the filter and cover. Ensure that filter cover is
securely installed. Replace the air filter regularly if it is
gray, torn, or the environment is dirty.
gray, torn, or the environment is dirty.
fuse
filter