Emerson Process Management CFA-3000 Manuel D’Utilisation

Page de 74
3
CFA­3000 ISE Analyzer
Instruction Manual
1.3   WARNING – Electrical Shock Hazard
 Installation of cable connections and servicing of this product require access to shock hazard voltage levels.
 Main power and relay contacts wired to separate power sources must be disconnected before servicing. 
 Non­metallic cable strain relief does not provide grounding between conduit connections! Use grounding 
type bushings and jumper wires. 
 Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA­ 70) and/or any
other applicable national or local codes. 
 Operate only with: 
1. Main console rear cover fastened. 
2. Left side power terminal cover closed. 
3. Keypad assembly latched closed. 
 Safety and performance require that this instrument be connected and properly grounded through a three­
wire power source. 
 Proper configuration and use is the responsibility of the user.
Mark
ID
Description
EN 61010­1
Safety of electrical equipment
EN 55011
Limits of RF equipment
EN 50082­1:92
Generic immunity from electrostatic  
discharge, RF & Fast 
Transient/Burst Immunity
EtL
9700516
ETL LISTED CONFORMS TO UL 
STD. 3101­1
EtL
ETL LISTED CERTIFIED 
TOCAN/CSA C22.2 No. 1010.1­92