Turn-On Checkout
Front Panel Description
1
2
3
N3280A   10V,   0.5A
Component Test DC Source
Figure 3-1. Front Panel, Overall View
Q
Q
Q
Q
 Line
      Switch
AC mains power switch.
R
R
R
R
 Unit
      Indicators
Unit indicators light to indicate the following operating conditions:
Power    The dc source is turned on.
Active    The dc source is addressed to talk or listen.
Error     There is a message in the SCPI error queue.
S
S
S
S
 Channel
      Indicators
Channel indicators light to indicate the following channel conditions:
On         The specified output channel is enabled.
Prot       The specified output channel has entered protection mode due to:
               Overtemperature,
               Overvoltage,
               Oscillation protect, or
               Power clear.
               Query the status registers of the affected channel to determine which