Agilent Technologies N3280A Manuale Utente

Pagina di 116
 
5 - Programming the DC Source
50
 Table 5-1. Bit Configurations of  Status Registers
 Bit
 Signal
 Meaning
 
 0
 1
 2
 3
 4
 5
 6
 
 CV
 CL+
 CL-
 CC
 VL+
 VL-
 OFF
 Operation Status Group
 The selected output is in constant voltage mode (applies only in voltage priority mode)
 The selected output is in positive current limit   (applies only in voltage priority mode)
 The selected output is in negative current limit  (applies only in voltage priority mode)
 The selected output is in constant current mode (applies only in current priority mode)
 The selected output is in positive voltage limit   (applies only in current priority mode)
 The selected output is in negative voltage limit  (applies only in current priority mode)
 The selected output is OFF
 
 0
 1
 2
 4
 10
 12
 14
 
 OV+
 OV-
 PCLR
 OT
 UNR
 OSC
 MeasOvld
 Questionable Status Group
 The positive overvoltage protection has tripped
 The negative overvoltage protection has tripped
 No communication with the selected output
 The overtemperature protection has tripped
 The output is unregulated
 The oscillation protection has tripped
 Output measurement exceeded capability of the range
 
 0
 2
 3
 4
 5
 7
 
 OPC
 QYE
 DDE
 EXE
 CME
 PON
 Standard Event Status Group
 Operation complete
 Query error
 Device-dependent error
 Execution error
 Command error
 Power-on
 
 2
 3
 4
 5
 6
 
 7
 
 WTG
 QUES
 MAV
 ESB
 MSS
 RQS
 OPER
 Status Byte and Service Request Enable Registers
 The unit is waiting for a trigger
 Questionable status summary bit
 Message Available summary bit
 Event Status Summary bit
 Master Status Summary bit
 Request Service bit
 Operation status summary bit
 
Operation Status Group
 
The Operation Status registers record signals that occur during normal operation.  As shown below, the
group consists of a Condition, PTR/NTR, Event, and Enable register. The outputs of the Operation Status
register group are logically-ORed into the OPERation summary bit (7) of the Status Byte register.
 Register
 Command
 Description
 Condition
 STAT:OPER:COND? (@<channel list>)
 A register that holds real-time status of the circuits
being monitored. It is a read-only register.
 PTR Filter
 STAT:OPER:PTR <n> (@<channel list>)
 A positive transition filter that functions as described
under STAT:OPER:NTR|PTR commands in chapter 6.
It is a read/write register.
 NTR Filter  STAT:OPER:NTR <n> (@<channel list>)
 A negative transition filter that functions as described
under STAT:OPER:NTR|PTR commands in chapter 6.
It is a read/write register.
 Event
 STAT:OPER:EVEN? (@<channel list>)
 A register that latches any condition that is passed
through the PTR or NTR filters.  It is a read-only
register that is cleared when read.
 Enable
 STAT:OPER:ENAB <n> (@<channel list>)
 A register that functions as a mask for enabling specific
bits from the Event register. It is a read/write register.