Agilent Technologies N3280A Manuale Utente

Pagina di 116
 
6 – Language Dictionary
54
SCPI Programming Commands - At a Glance
Table 6-1. Subsystem Commands Syntax
ABORt
Resets the trigger system to the Idle state
CALibrate
    :CURRent
         [:LEVEL] (@channel)
Calibrate output current and low current measurement range
         :LIMit
              [:POSitive] (@channel)
Calibrate positive current limit
              :NEGative (@channel)
Calibrate negative current limit
         :MEASure <max_val>, (@channel)
Calibrate high and medium current measurement range
    :DATA <n>
Enters the calibration value
    :DATE <date>
Sets the calibration date
    :LEVel <level>
Advance to next calibration step (P1 | P2)
    :PASSword <n>
Set numeric calibration password
    :SAVE
Save new cal constants in non-volatile memory
    :STATE <bool> [,<n>]
Enable or disable calibration mode
    :VOLTage (@channel)
Calibrate output voltage and voltage measurement range
INITiate
    [:IMMediate]
          :NAME <name>
Enable the named trigger system (TRANsient | ACQuire)
FETCh
    :ARRay
          :CURRent [:DC]? (@list)
Returns the digitized instantaneous current
          :VOLTage [:DC]? (@list)
Returns the digitized instantaneous voltage
    [:SCALar]
          :CURRent [:DC]? (@list)
Returns output current dc measurement
          :VOLTage [:DC]? (@list)
Returns output voltage dc measurement
MEASure
    :ARRay
          :CURRent [:DC]? [max_val,] (@list)
Digitizes and returns the instantaneous output current
          :VOLTage [:DC]? (@list)
Digitizes and returns the instantaneous output voltage
    [:SCALar]
          :CURRent [:DC]? [max_val,] (@list)
Digitizes and returns average (dc) output current
          :VOLTage [:DC]? (@list)
Digitizes and returns average (dc) output voltage
OUTPut
    [:STATe] <bool>,  (@list)
Enables/disables the selected dc source output
    :OSCProtect
          [:STATe] <bool>,  (@list)
Enables/disables oscillation protection on the selected output
    :PROTection
          :CLEar  (@list)
Reset latched protection
SENSe
    :CURRent[:DC]
          :RANGe [:UPPer] <max_val>, (@list)
Selects the current measurement range
    :FUNCtion <function>, (@list)
Configures the measurement sensor (
"VOLTage" | "CURRent"
 )
    :SWEep
          :NPLCycles <n>
Sets the number of ac power line cycles
          :OFFSet
                :POINts <n>
Defines the trigger offset in the measurement sweep
          :POINts <n>
Defines the number of data points in the measurement
          :TINTerval <n>
Sets the measurement sample interval
    :WINDow [:TYPE] <type>
Sets measurement window function 
(HANNing | RECTangular)