Patient Safety Tests
4-42
Testing the Monitor
Enclosure
Leakage
Current
This test is compliance with EN60601-1 (Enclosure Leakage Current). Test at 110% of the
nominal line voltage.
nominal line voltage.
The analyzer leakage current indication must not exceed the values listed below.
Table 4-5. Enclosure Leakage Current
EN60601-1 (1990 + A1, A2, A11, A12, A13) and UL2601 (2
nd
Ed. 1997) US Deviations
What to record on your service record:
S(3): P or S(3): F
Step
Action
1
Connect the AC mains power cord to the analyzer as recommended by the
analyzer operating instructions.
analyzer operating instructions.
2
Using the appropriate test cable, connect the analyzer to either of the screws on
the back of the monitor, next to the handle.
the back of the monitor, next to the handle.
3
Turn the monitor on.
4
Perform the test as recommended by the analyzer operating instructions.
Test Condition
Polarity
Allowable Leakage Current
Normal
Normal
100 µA
Reversed
100 µA
S.F.C
1
Open
Protective Earth
Normal
300 µA
Reversed
300 µA