Monitoring EEG
The Electroencephalograph (EEG) module monitors the patient’s cerebral function by measuring the 
electrical activity of the brain. It provides the monitor with two channels of realtime EEG waves, EEG 
trend information in the form of Compressed Spectral Arrays (CSA), and up to eight of the following 
numerics: 
Spectral Edge Frequency (SEF): The SEF is the frequency below which a defined percentage of the 
Total Power lies. The percentage is set in Configuration Mode.
Mean Dominant Frequency (MDF): The MDF is the mean value of the frequency which dominates 
the measured EEG.
Peak Power Frequency (PPF): The PPF is the frequency with the highest measured amplitude.
Total Power (TP): The TP numeric indicates the power in the measured frequency band.
Percentage of total power in each frequency band:
Alpha
 waves (8 to 13 Hz)
Beta
 waves (13 to 30 Hz)
Theta
 waves (4 to 8 Hz)
Delta
 waves (0.5 to 4 Hz).