Agilent Technologies 8935 series e6380a Manuale Utente

Pagina di 320
Chapter 5
139
CDMA Measurements - Screens and Control Fields
CDMA Measurements
Code Domain Measurements – IS-95
If you have selected IS-95 only in the INSTRUMENT CONFIGURE 
screen, the Code Domain screen will appear as described in this section, 
with the Walsh Code ordering following the Hadamard paradigm, with 
64 channels, total. These channels are referred to as “I” channels.
If you wish to make use of the extended Walsh ordering - Bit Reverse - 
as described by the IS-2000 standard, see See “Code Domain 
Measurements – IS-2000” on page 142.
The following measurements are displayed on the CODE DOM (code 
domain analyzer) screen: power, fast power, timing and phase. 
Measurements are selected from the 
Measurement
 field on the 
Main
 
controls menu.
Time offset (Tm Ofs) and frequency error (Freq Err) are always 
displayed when any code domain measurement is made. Carrier 
feedthrough (CarFT) is also displayed except when fast power is 
measured.
Figure 5-5 
Code Domain Power
Code Domain Power, IS-95
The Test Set provides two code domain power measurements.
• Absolute code domain power displays the power in each of the 64 
Walsh channels, relative to the total power inside a 1.23 MHz 
bandwidth centered at the tune frequency. The 64 Walsh channels (0 
through 63) are represented by a vertical bar on the analyzer’s 
display. To measure absolute code domain power the 
Ch pwr unit
 
field on the 
CD Setup
 menu must be set to 
Abs
 (see 
). Use 
the 
Marker
 controls to move the marker to the Walsh channel (
Walsh 
Chan
) you want to measure (see 
). 
Code D om ain
Frequency D om ain
0
1
2
3
4
5
6
7
8
9
 
32
 
 
 
63
User
1
User
3
User
2
Walsh Code
Pilot Paging
Synch
U s e r  # 2
U s e r  # 1
S yn ch
P a g in g
P ilo t
1.2288 MHz
fre q
O th e r  u se rs  &   n o ise
Frequency Domain
Code Domain