Agilent Technologies 8935 series e6380a Manuale Utente
Chapter 5
173
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
Time Offset
This field allows you to manually enter a time offset. To obtain a value
to enter in this field you must make at least one of the following
measurements.
to enter in this field you must make at least one of the following
measurements.
• Power
• Timing
• Phase
• Rho
After making the measurement, access the
FP Setup
menu on the
CODE DOM screen, and enter the value obtained from the
Time Ofs
field or, for rho measurements, the
Time Offset
measurement field on
the CDMA ANALYZER screen.
An entry into the
Time Offset
field can also be made automatically,
using the
Ofs Trnsfer
field. See
.
When measuring fast power, the code domain analyzer uses the value
in the
in the
Time Offset
field instead of re-measuring the time offset
parameters each time a code domain measurement is made.
Screens on which this field is present
CODE DOM (Code Domain Analyzer, IS-95 only): FP setup controls
menu
menu
Time Ofs
This measurement field displays the time offset when a rho
measurement is made. See
measurement is made. See
for more
information about this measurement.
Screens on which this field is present
CODE DOM (Code Domain Analyzer).