Chapter 6
219
Analog Measurements - Screens and Control Fields
Control Fields for Analog Measurements
Trig-Delay
The trigger delay is used to specify the time relationship between the 
trigger and displayed signal. 
Positive
 values delay the measurement trigger by a specific period. 
The delayed trigger point is the left edge of the screen. 
Negative
 values perform a pre-trigger function, displaying a section 
of the waveform before the trigger point. The trigger point is 
indicated by small pointers that appear at the top and bottom of the 
screen. 
Operating Considerations
Negative Value   The maximum negative delay cannot exceed ten 
divisions of the current time-per-division setting. For example; if the 
Time/Div
 field is set to 1 ms, the maximum allowed negative delay is 
10 ms. Larger negative numbers cause an 
Excessive negative 
Trig-Delay will be truncated.
 message. 
Positive Values   For 
Time/Div
 settings of 50 
µ
s/div and smaller, the 
maximum delay is 400 ms. 
For 
Time/Div
 settings of 100 
µ
s/div and larger, the maximum delay is 
3300 ms. 
Resolution
 For delays of 400 ms and less, the resolution is 6.4 ms. For delays 
greater than 400 ms, the resolution is 51.2 ms. All entries are rounded 
to the nearest multiple of 6.4 ms or 51.2 ms (depending on the delay 
value). 
Screens on which this field is displayed
SCOPE: trigger controls menu