Chapter 8
275
Signaling Encoder and Decoder - Screens and Control Fields
Control Fields for the Signaling Encoder and Decoder
Send Mode
This field selects the mode used when
Send
is selected to output the
message.
•
Single
outputs the entire message once.
•
Burst
outputs the Message the number of times specified in the
Bursts
field.
•
Cont
causes the message to be output continuously until
Stop
is
selected.
•
Step
is not used in the AMPS-TACS mode.
Screens Where Field is Present
SIGNALING ENCODER (Modes: AMPS-TACS, NAMP-NTAC, DTMF)
Sine Units
This field specifies whether the signal’s output is in units of rms or
peak. This field is only present when the
peak. This field is only present when the
AFGen2 To
field is set to
Audio
Out
.
Screens Where Field is Present
SIGNALING ENCODER (Modes: Func Gen)
Single/Cont
This field specifies how long you want the analyzer to decode incoming
signals:
signals:
•
Single
tells the analyzer to display the information received during
one
Gate Time
. Measurements are displayed until
Arm Meas
is
selected again.
•
Cont
is used to automatically re-arm the analyzer and display the
measurements on a continual basis until
Single
is selected.
Previous measurement results are over-written by subsequent
measurements.
measurements.
Operating Considerations
To dis-arm the decoder in single mode, select the
Stop Meas
field. The
Stop Meas function is disabled when
Cont
is selected.
Screens Where Field is Present
SIGNALING DECODER (All Modes)